| Test voltage | ||||||
|---|---|---|---|---|---|---|
| Measurement current | 1 mA (Test voltage 250 V to 1.00 kV), 0.5 mA (Test voltage 1.10 kV to 2.50 kV) 0.25 mA (Test voltage 2.60 kV to 5.00 kV), Short-circuit current: 2 mA or less | |||||
| Resistance range | 10 MΩ to 10 TΩ, 7 ranges (auto range) | |||||
| Accuracy | ±5% rdg. ±5 dgt. Up to [Test voltage (setting value)/Resistance measurable at 100 nA] ±20% rdg. ±5 dgt. [Test voltage (setting value)/Resistance measurable at 100 nA] to [Test voltage (setting value)/Resistance measurable at 1 nA] or 500 GΩ ±30% rdg. ±50 dgt. [Test voltage (setting value)/Resistance measurable at 1 nA] or 501 GΩ to 9.99 TΩ |
|||||
| Leakage current measurement | 1.00 nA to 1.20 mA, 6 ranges (current measurement that occurs when test voltage is generated) Accuracy ±2.5% rdg. ±5 dgt. (1 mA range); refer to complete catalog for other ranges |
|||||
| Voltage measurement | ±50 V to ±1.00 kV DC, 50 V to 750 V AC (50/60 Hz), Accuracy: ±5 % rdg. ±5 dgt. , Input resistance: Approx. 10 MΩ |
|||||
| Temperature measurement | -10.0 ˚C to 70.0 ˚C, 3 ranges (used with optional sensor) Accuracy ±1.0 ˚C (0.0 ˚C to 40.0 ˚C); refer to complete catalog for other ranges |
|||||
| Other functions | Insulation Diagnosis (Temperature compensation, PI/DAR display, Step voltage test), Data memory, Communication (USB 2.0, PC application software), auto discharge, hot conductor warning indication, etc. | |||||
| Display | Digital LCD, max. 999 dgt. with backlight, Bar graph display | |||||
| Power supply | LR6 (AA) alkaline batteries ×6, Battery pack 9459, or AC adapter 9753 or 9418-15 (100 – 240 VAC) Continuous use : [LR6] 5 hr, [9459] 9 hr, (Occur 5 kV, +/- open terminal) |
|||||
| Dimensions and mass | 260 mm (10.24 in)W × 250.6 mm (9.87 in)H × 119.5 mm (4.70 in)D, 2.8 kg (98.8 oz) | |||||